Sensor Head Installation (Side Mount Configuration)
OPS
Series Encoders
2.
Verify Sensor Mounting Surface Height
Refer to the OPS interface drawing for detailed dimensions.
Use Z-Height spacer shim to verify that the Z-Height distance between
the bottom surface of the sensor and the top of the scale is as follows:
Tape Scales:
Use red spacer (1.00mm) to set the proper Z-Height for
PurePrecision Marker Tape II and Laser Tape II Scales
Glass Scales:
Use blue spacer (0.83mm) to set the proper Z-Height for
PurePrecision Performance and Value Linear Glass Scales
See step 3 below for installation steps.
3.
Installation Steps
3.2 Place shim between the bottom of the sensor head
and the top of the scale and press down in the Z-axis.
Shim can be inserted either parallel or perpendicular
to the scale. The shim features cut-outs to allow
signal level to be monitored during sensor mounting.
Z
Y
X
q
z
Axis Diagram
3.1 Loosen the sensor in the Z-axis
3.4 Carefully remove the shim by rotating it off the
scale with the shim’s handle.
3.3 Tighten the sensor in the Z-axis.
Rotational
movement to
remove shim.
Gently push the
sensor and
mounting against
the top of the
tape scale in the
z-axis.
1.
Install the Sensor
Use a wrench and M3 screws to install the sensor. Refer to interface drawing to make sure sensor is oriented
properly with reference to the scale.